Technical data:
• pH Range: 0.00 – 10.50
• Temperature Range: 10 – 80 Celsius degree
• Reference Element: Ag/AgCl, saturated KCl
• Internal Electric Resistance: maximum 250 Mohmi at
20 Celsius Degree
• Null point at pH 7.00 at 20 Celsius Degree
• Asymetry Potential: maximum 35 mV at 20 Celsius Degree
• Junction: ceramic
• Cable: coaxial 50 ohmi
(- for laboratory use: 1 m coaxial
cable
- for process use: maximum 10 meters coaxial cable is recommended;
over 10 meters a high impedance signal transmission is required)
• Plug: BNC
• Body glass diameter: 11.5 mm; 12.5 mm |